Fails TPS Reset when booting up to Applications or when Resetting TPS due to XGD Amplifier.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2256925 StartupCondition error on XGD StartupCondition resulted in TPS reset failure See Long Description 2256925 SLMAINMR LxMxConsumer services/initialization/ServerInitManager 767 Server Name: SC StartupCondition error on XGD StartupCondition resulted in TPS reset failure.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
Area
MR