Line (zipper) artefact in Whole Body scan was caused by the Roller of Table.

Symptom

Line (zipper) artefact in Whole Body scan was caused by the Roller of Table. It makes the table movement uneven during whole body scan. A Co57 flood source was put under the line phantom for a whole body scan for testing. A zipper artefact was seen in the image.

Applies to

Models
Infinia
Product
Infinia II Infinia II Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software) Infinia I Infinia I Acquisition System
Component
ASM001060 Roller Assembly Infinia Table Top ASM001060 Roller Assembly Infinia Table Top *Patient Table *Patient Table *Patient Table
Area
Nuclear Medicine