Pular para o conteúdo principal
GE Healthcare · MRSigna2252306

TR Driver Fault Detected Short or Open Circuit Source of Fault is Head TR Due to 3T Corona Head TR Switch.

Symptom

Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.

Error message

2252306 TR Driver Fault Detected Short Circuit Source of Fault is Head TR See Long Description for Extended Error. 2252306 ERROR: EM_DCB_TR_SHORT_CKT TR Driver Fault Detected - Short Circuit. Source of Fault is %b. Value: %d mA Limit: %d mA Extended Error: If this error happens repeatedly then check for a short-circuited PIN diode or other short-circuit condition somewhere in the failing transmit chain. Use TR DD MC System Path diagnostic or Gather Load Data and verify that the short circuit values are less than 30mA. If error happens intermittently and cannot be traced to failed hardware then it may be due to gradient noise in the system.

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2362992 3T CORONA HEAD ASSEMBLY 2362992 3T CORONA HEAD ASSEMBLY. 3T Head T/R Switch.
Area
MR

Full cause & step-by-step resolution