Pular para o conteúdo principal
GE Healthcare · CTLightSpeed

LightSpeed 4.X - GRE

Symptom

HP Host computer intermittently locks up or will not boot up to applications HP 8000 or 8200 computer intermittently locks up or will not boot up to applications

Error message

IN /var/log/messages: kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error See long description for details IN /var/log/messages: uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error Open unix shell and login as root cd /var/log cat messages | grep Error Look for any errors similar to these disk drive errors shown below: Mar 21 15:49:38 uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error Mar 21 15:49:43 uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error Mar 21 15:49:48 uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error Mar 21 15:49:53 uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error Mar 21 15:49:59 uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error Mar 21 15:50:04 uact03 kernel: Info fld=0x3c27e5d, Current sdc: sense key Medium Error

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5117866-5 73 GB HDD used in HPxw8200 GRE. See Long Discription for Details. Part #: 5117866-5 Part Type: GP CONSUMABLES Part Weight: -139998.60 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73 GB HDD for Lighting System Controller 5117866-2 FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues