Unable to Scan with The GP is reporting a Gradient Amplifier Axis Internal Wiring Fault due to HFA Control Board.Replace…

Unable to Scan with The GP is reporting a Gradient Amplifier Axis Internal Wiring Fault due to HFA Control Board.Replace HFA CONTROL BOARD ASSEMBLY wqr.Signa HDxT 2244570 The gradient amplifier failed…

Symptom

Will Not Scan with Gradient Errors.

Error message

2244570 The gradient amplifier failed to go to Ready when commanded. 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded.-> 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded.

2244847 The GP reported a Gradient Amplifier % Axis Internal Wiring Fault or Control Undervoltage Error: 2244847 Host: SEMRADMR2 Process: NSP File: SCP:Grad Line: 2615 Exception Class: Unknown Severity: Unknown The GP reported a Gradient Amplifier X Axis Internal Wiring Fault or Control Undervoltage.

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2396050-2 HFA CONTROL BOARD ASSEMBLY
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.