Pular para o conteúdo principal
GE Healthcare · MRExcite22521062223849

TPS fails reset fails due to failed MGD power supply.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2252106 Failed to determine hardware state for AGP board within 20 seconds Error: 2252106 Host: rmns Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for AGP board within 20 seconds The board may be busy loading vxWorks or may be hung.

2223849 AGP, SCP, APS EPROM power-up boot failed

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
5181495-3 400W Quad Output 6Ux8HP cPCI Power Supply This is a replacement for EOPL 2294300-8
Area
MR