GE Healthcare · MRExcite22521062223849
TPS fails reset fails due to failed MGD power supply.
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
2252106 Failed to determine hardware state for AGP board within 20 seconds Error: 2252106 Host: rmns Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for AGP board within 20 seconds The board may be busy loading vxWorks or may be hung. 2223849 AGP, SCP, APS EPROM power-up boot failed
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 5181495-3 400W Quad Output 6Ux8HP cPCI Power Supply This is a replacement for EOPL 2294300-8
- Area
- MR
Full cause & step-by-step resolution