Pular para o conteúdo principal
GE Healthcare · MRSigna22245894062796

TG Reached the High Limit Due to SRFD2 RF Amplifier.

Symptom

Auto Prescan Fails with TG Reached the High Limit See error in Long Description 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan.

Error message

2224589 TG reached the high limit. See Long Description Suite: MR01 Host: OC0 Proc: MSG_HANDLR0 Error: 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan. Note: On later systems, this same error code has been slightly reworded to read as follows: Thu Jun 12 11:06:12 2008 Error: 2224589 Host: MENEMR Process: NSP File: TPS:aps_exec Line: 4062796 TG reached the high limit. TG has been set to 190. Check patient positioning and if desired try Auto Prescan again or use Manual Prescan.

Applies to

Models
Signa
Product
Signa HDxT Apps. For use with HD16.0_VO1 and HD16.0_VO2 version software only. For use with HD16.0_VO1 and HD16.0_VO2 version software only.
Software
HD16 Includes HD16.0_VO1 and HD16.0_VO2 version software and applies to 1.5T and 3.0T "Signa HDxT Apps" systems only. Includes HD16.0_VO1 and HD16.0_VO2 version software and applies to 1.5T and 3.0T "Signa HDxT Apps" systems only.
Component
SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.