GE Healthcare · MRDiscovery2274021

Cannot scan with Error 2274021 Coil Fault - Short Circuit A multicoil bias fault was detected due to coil.Replace Coil w…

Cannot scan with Error 2274021 Coil Fault - Short Circuit A multicoil bias fault was detected due to coil.Replace Coil wqr.

Symptom

Cannot scan with error 2274021 Coil Fault Short Circuit multicoil bias fault was detected see long Connector: Port 1 (P1) Channels faulted: 0x10 First channel: 5 of 16, counting from 1 Fault value: -2246 mV, Channel: 5->

Error message

2274021 Coil Fault - Short Circuit A multicoil bias fault was detected See Long Description Error: 2274021 Host: WBH_RAD_MR11 Process: NSP File: SCP:RfHubCanHWO::RfBias Line: 4080 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HNS CTL123 Scan: 4637/1 Coil Fault - Short Circuit A multicoil bias fault was detected. Connector: Port 2 (P2) Channels faulted: 0x100 First channel: 9 of 32, counting from 1 Fault value: -4777 mV, Channel: 9

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5271265-2 3.0T HD 8Ch Shoulder Coil Assembly 5271265-2 3.0T HD 8Ch Shoulder Coil Assembly
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.