GE Healthcare · Nuclear MedicineInfinia

Intermittent gantry errors #47, #103, #120

Symptom

Intermittent gantry errors #47, #103, #120

Error message

Error # 47 (Error 47) ACQUISITION HALTS PSD TEST FAILURE Corrective Action: Replace the Collimators. If the problem persists, replace the Collimator ID board on both collimators Check Connectors P7 and P405/P1205 on HCIC board of Head #1 and Head #2, respectively. Check interior Head cable (CBL001314) between HCIC P7 and the CBCIC J4 for both Heads. Check the cnnection of Connector J1 on the CBCIC board for both Heads. Replace the FOXPOP board.

Error #103 (Error 103) GANTRY_ERROR_ROD_PARK Axis is in motion while cart is in place GANTRY_ERROR_ROD_PARK ERROR 103 Axis is in motion while cart is in place

Error # 120 (Error 120) Collimator unlocked axis = $ Where $ refers to a specific axis Axis # Axis Name 0 Lateral 1 Radial #1 2 Radial #2 3 N/A 4 N/A 5 Swivel #1 6 Swivel #2 7 Rotation 8 Linear 9 Height

Applies to

Models
Infinia
Product
Infinia II Infinia II Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software) Infinia I Infinia I Acquisition System
Component
*Gantry Gantry Gantry
Area
Nuclear Medicine

Full cause & step-by-step resolution