Pular para o conteúdo principal
GE Healthcare · CTLightSpeed20010901220050200262920704

LightSpeed 4.X - GRE

Symptom

Unable to scan axial or helical; system cannot allocate image space as one chunk. Unable to scan axial or helical; system cannot allocate image space as one chunk

Error message

Error: 200109012 Image Allocation Failed. Cannot build scan request. See Long Description for complete error Wed Mar 2 18:39:33 2005 Error: 200109012 Host: ct01 Process: scanRx File: UIRxToScanner.cxx Line: 1307 Function: Data Acquisition : OC Processing Image Allocation Failed. Cannot build scan request.

Error: 200502002 Cannot allocate image space as one chunk. See long description for error example Error: 200502002 Cannot allocate image space as one chunk Wed Mar 2 18:39:33 2005 Error: 200502002 Host: ct01 Process: scanRx File: srx_sma_manage_image_space.c Line: 682 Function: No system function reported Cannot allocate image space as one chunk. Total space of 62920704 (bytes), 2 paths returned.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
2.4.2_H2.4M5 for LightSpeed 2.X (Plus, QX/I) with SGI Octane Consoles running the IRIX OS 2.4.2_H2.4M5 for LightSpeed Plus 2.x with SGI Octane Consoles running the IRIX OSPart # Desc.2395178 - LS2002 M5 SDAS Software Collector 2395182 - LS2002 M5 Application Collector 2395160 - Apps Disk 1 2395161 - Apps Disk 2 2356012 - Irix OS Collector 2356015 - IRIX OS Disk 1 2356016 - IRIX OS Disk 2 2356017 - IRIX OS Disk 35113667 - LS2002 M5 Patch C.12395172 - LightSpeed Class C - 1104I.2_H4.0M5CLASSC2395173 - LightSpeed Class M - 1204I.2_H4.0M5CLASSM 2.4.2_H2.4M5 for LightSpeed Plus 2.x with SGI Octane Consoles running the IRIX OS FMI 25332 - LS2002 M5 Software Upgrade The purpose of this FMI is to upgrade LightSpeed IRIX-based CT scanners to the latest version software. This software includes safety and quality fixes. Safety (PSR / FPR) fixes include: • PSR 1286436 -- Auto mA table values, • PSR 1286629 -- Auto mA noise index, • PSR 13000489 -- Scan abort with prep delay, and • PSR 1286829 -- Wrong Patient Information. Quality fixes include: • Scout not displayed, • ring artifact, • remote tilt, • align light, • Autolink page slow, • Batch Install Image not displayed, • Image out of order, • Slow image-biopsy, • Transfer to Pacs Error, • Service Desktop Error, • autoMA-concerns (table), • Modem, • autoMA-noise (GUI), • death of stclf, • axial not ready, • Smart Prep fails to start, • .rhosts file corrupted, • scan data missing in retro, • Square VP not saved, • AutoArchive, • peripheral gate hang, • ali pacs (mekesson), • Can't recover MOD images, • not accept image - Phillips, • Show Local lines delay, and • Head IQ. New delivered features are: • AutomA GUI, • Repeat Series, • Re-Start / Shutdown Procedure, and • Anonymous Scan Data Save. Furnished Materials Qty Part number Description 1 2399103-100 FMI document 1 2395178 LS2002 M5 SDAS Software Collector containing: 1 2395182 LS2002 M5 Application CD Collector (3 CDs) 1 2356012 LightSpeed OS Software Collector (3 CDs) 1 46-302200P2 Product Locator 1 2348269-100 LS2002 Software Install Document 1 2243321-100 LightSpeed 2X Apps Tips and Workarounds 1 2369740-200 CT Lightspeed Series Learning and Reference Guide Tools Required Qty Part # Description 1 2395172 LightSpeed Class C Software (Tab Distribution) 1 2395173 LightSpeed Class M Software (Tab Distribution) 1 2347142 Variviewer Software Option (Customer Site) 1 2356009 Auto mA Software Option (Customer Site) 1 2345411 Power 440 Software Option (Customer Site) - - Customer Purchased Options (Customer Site) Synonyms: LSPlus, B7850JE, B7850JF, B7850JL Gantry: 2260619, 2260619-3, B7850NG Console: 2266832 2266832-2 2304732, 2304732-2, B7850CD B7850NC B7850ND
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.