Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260007370

LightSpeed 4.X - Octane

Symptom

Customer sees a Pop-up window indicating the DAS temperatures are low. This does not prevent scanning only a warning. Customer sees a Pop-up window indicating the DAS temperatures are low. This does not prevent scanning only a warning

Error message

Error: 260007370, DAS Temperature is LOW. See Long Description for complete error Mon Feb 20 07:43:59 2006 Error: 260007370 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 490 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:43:59:965 Log Series: 32 DAS Temperature is LOW. Lowest temperature was seen is on Converter Board 31. Measured: 10.5 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. See Following Error Message for Converter Card List

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.