Equipment Room temperature exceeds GE specification window of 59-89.6°F (15-32 °C).

Symptom

Scan stops with gradient error 2244897.

Error message

2244897 The GP is reporting a % Axis OverTemperature Fault See Long Description System will wait 5 minutes before attempting recovery. Extended Error: Check that the cabinet fan is functioning properly, rear door is closed and sealed, and that no filler panels have been removed from the front of the cabinet. Proper cooling requires a vaccum to pull air through the Gradient Amplifiers, Power Supply, and GP. If fault persists on specific PSDs write a PQR. If fault persists on many or all PSDs, replace Gradient Amplifier.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Area
MR