The signal is too small due to damaged quad knee coil cables.

Symptom

Auto Prescan Fails With Knee Coil

Error message

2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2383613-3 Medical Advances Quad Knee Coil Cable Assembly MED ADV QUAD KNEE CABLE.
Area
MR

Full cause & step-by-step resolution