Pular para o conteúdo principal
GE Healthcare · CTLightSpeed244518178588183190

LightSpeed 4.X - GRE

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 244518 Scan hardware reset failed See Long Description for complete error Sun Sep 26 08:30:10 2004 Error: 244518 ctge04 fwmgr HwManager.c 1536 Scanning hardware reset failed .

Error: 178588 Failed to download charaterization file 4-fw/config/Ro tor.cfg Tue Oct 11 01:18:55 2005 Error: 178588 Host: ct02 Process: fwmgr File: Characterization.c Line: 280 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to download charaterization file 4-fw/config/Ro tor.cfg Wed Apr 1 13:44:43 2009 Error: 178588 Host: sea_rad_ct1 Process: fwmgr File: Characterization.c Line: 281 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to download charaterization file 13-fw/char/CalPts.char

Error: 183190 Tube ID parameter in Rotor Characterization file does not agree with tube.Tube ID parameter in Rotor Characterization file = 8. Actual tube ID =0. Rotor Characterization file rejected. See the Long Description for the complete error code Tue Oct 11 01:19:16 2005 Error: 183190 Host: OBC Process: Generator File: gen_in_mail.c Line: 1521 Function: XRAY GENERATION : Tube Identifiers Scan: 0/0/0 Type: None/Unknown E

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-288512G1 GENTRY I/O BOARD See LD for details Part #: 46-288512G1 Part Type: GP REPAIRABLES Part Weight: 1.36 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: GENTRY I/O BOARD FRU: Yes Status: Orderable Comments: FLS - CLASS 1 FRU CT PULL PART
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.