Pular para o conteúdo principal
GE Healthcare · MRExcite2248673

MDS Link Errors Due to Gradient Processor

Symptom

TPS or MGD Reset Fails With MDS Link Errors

Error message

2248673 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. 2248673 Host: lxmr Process: NSP File: SCP:MdsLinkHWO Line: 1713 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. Check gradient cabinet power, fiber optic connections and signal level of fiber optic link. Run MDS fiber optic link diagnostics.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2248992-2 Gradient Processor GP1
Area
MR

Full cause & step-by-step resolution