Defective Master Exciter2 board.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2247159 Sine Wave Table on RIF DIF1 Board Did Not Load Correctly See Long Description Fri Jan 25 09:23:14 2008 Error: 2247159 Host: FHMR01 Process: diagCoord File: Irf.cpp Line: 412 Diagnostic: MUX BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 51174 Object: APS Test Task The Sine Wave Table on the DIF board in RRF chassis (front, slot 10) did not load correctly. This will prevent proper Exciter output. Try TPS_reset. Run RRF diagnostics. Reported from Irf::reset_rrf (0) (0) (150)->

2247157 The Master Exciter in the RRF chassis (rear, slot 7) was not detected. Error: 2247157 Host: ge3T Process: diagCoord File: Rrf.cpp Line: 249 Task ID: 0x3f1fdf8 Time: 999 Object: APS The Master Exciter in the RRF chassis (rear, slot 7) was not detected. The system will not function without this board.-> Error: 2247157 Host: ge3T Process: diagCoord File: Rrf.cpp Line: 249 Task ID: 0x3f1fdf8 Time: 999 Object: APS The Master Exciter in the RRF chassis (rear, slot 7) was not detected. The system will not function without this board. Extended Error: The master exciter provides the main clock for the RRF chassis and the IRF board. Verify that the board is powered and that its LO DETECT LED is

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.