GE Healthcare · MRSigna2247159511742247157224716051176
Defective Master Exciter2 board.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2247159 Sine Wave Table on RIF DIF1 Board Did Not Load Correctly See Long Description Fri Jan 25 09:23:14 2008 Error: 2247159 Host: FHMR01 Process: diagCoord File: Irf.cpp Line: 412 Diagnostic: MUX BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 51174 Object: APS Test Task The Sine Wave Table on the DIF board in RRF chassis (front, slot 10) did not load correctly. This will prevent proper Exciter output. Try TPS_reset. Run RRF diagnostics. Reported from Irf::reset_rrf (0) (0) (150)-> 2247157 The Master Exciter in the RRF chassis (rear, slot 7) was not detected. Error: 2247157 Host: ge3T Process: diagCoord File: Rrf.cpp Line: 249 Task ID: 0x3f1fdf8 Time: 999 Object: APS The Master Exciter in the RRF chassis (rear, slot 7) was not detected. The system will not function without this board.-> Error: 2247157 Host: ge3T Process: diagCoord File: Rrf.cpp Line: 249 Task ID: 0x3f1fdf8 Time: 999 Object: APS The Master Exciter in the RRF chassis (rear, slot 7) was not detected. The system will not function without this board. Extended Error: The master exciter provides the main clock for the RRF chassis and the IRF board. Verify that the board is powered and that its LO DETECT LED is
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.