Pular para o conteúdo principal
GE Healthcare · CT18392128848503891265527513081394183905

Lightspeed 1.X - Octane

Symptom

"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors

Error message

Error: 183921 Axial conditions at FAULT: See Long Description for complete error Mon Oct 17 08:26:48 2005 Error: 183921 Suite: SMCT Host: STC Process: Axial File: ax_error.c Method: No Method Line: 1114 Function: AXIAL CONTROL : Axial Servo Control Loop Scan: 28848/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 08:26:47:775 Log Series: 141 Axial conditions at FAULT: velocity_fb = 2662 position_fb = 50389 Formatted data: 8100 100 8 7 1 0 1 1 101 1 0 2662 -12655275 13081394 2741 2582 134 133 0x019e38

Error: 183905 Axial MOTION with brake on (moved at least 1 degree). See long description Wed Dec 30 08:09:21 2009 Error: 183905 Host: STC Process: Axial File: ax_error.c Line: 1114 Function: AXIAL CONTROL : Axial Servo Control Loop Scan: 17063/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 08:09:21:685 Log Series: 51 Axial MOTION with brake on (moved at least 1 degree). position: commanded = 0.0 degs measured = 1.1 degs HW Error: Retry Rx to verify probable Gantry Balance or Brake or Encoder Hardware problem

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.