Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260101023288612601010125006026010107150018

LightSpeed RT16 & RT16 Xtra and Vision

Symptom

Attention: The Scanning Hardware is not available. Please see long description. Attention: The Scanning Hardware is not available, please wait for the hardware reset successfull status posted in the feature status area on the right monitor. If the hardware reset fails, then reset the scan hardware from the Service Desktop System Resets button.

Error message

Error: 260101023 The CCB has notified the system of an error condition which prevents scanning Thu Nov 20 10:38:27 2008 Error: 260101023 Host: orp Process: orp_subsys File: rcib_erlog.c Line: 463 Thu Nov 20 10:38:26 2008 Host : Collimator Ermes # : 260101023 Exception Class : Abort Severity : Primary File : rcib_erlog.c Line# : 463 Function : No System Function Reported Scan Type : Scout Scan: 28861/1/1 The CCB has notified the system of an error condition which prevents scanning. See log for further information

Error: 260101012 The Rotating Controller Interface Bus (RCIB) fault line has been asserted. See LD for more details Mon Apr 10 07:20:41 2006 Error: 260101012 Host: orp Process: orp subsys Mon Apr 10 07:20:40 2006 Host : On-Board Rotating Processor Ermes # : 260101012 Exception Class : Abort Severity : Primary File : rcib fault state.c Line# : 139 Function : No System Function Reported Scan Type : Axial Scan: 50060/6/11 The Rotating Controller Interface Bus (RCIB) fault line has been asserted.None of the sub-systems have reported a problem. Check nodes or fault line for failure.

Error: 260101071 The DCB has reset unexpectedly See Long Description for full error Error: 2601

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2404393 GDAS-16 CONVERTOR CARD, replaces 2364354. See Long Description for details. 2404393 GDAS-16 CONVERTER CARD, replaces 2364354 Part #: 2404393 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: 16 SLICE GDAS CONVERTER CARD FRU, A CONVERTER CARD IN AN ESD BAG IN A BOX WITH A PART NUMBER LABEL FRU: Yes Status: Orderable Comments: FLS - 16 SLICE GDAS CONVERTER CARD USED ON: LIGHTSPEED PRO 16 FRU GDAS ADDED PER GDAS RSPL MS13452 040315 RISK: SITE DOWN FREQUENCY: HIGH NOTE: THERE ARE 96 OF THESE CARDS PER SYSTEM, BUT SHOULD NOT NEED TO REPLACE ALL AT ONETIME, MS13452 THIS PART CONTAINS ONE GDAS CONVERTER CARD 2364354 WITH FRU PACKAGING, SUPPLIED BY GEMS IT. MS13452 PDR 8471.80.9000 IM05 UPDATE AWAITING MM (212001261 4/1/06) For 4/8 slice G Das which uses a thick connector must order 2364354-2
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.