GE Healthcare · CTLightSpeed26011812050103260118185
LightSpeed 5.X (60kW)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118120 (20-0255) Filament Open circuit Ermes: 260118120 (20-0255) Filament Open circuit Wed Mar 21 08:25:45 2007 Error: 260118120 Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Wed Mar 21 08:25:44 2007 Host : Jedi Ermes # : 260118120 Exception Class : Secondary Severity : Pri/Soft File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Axial Scan: 50103/8/6 (20-0255)Filament open circuit(inverter1)(<3 times). ^C Z6,Z7= 97h,0h Error: 260118185 (50-0204) Filament permanent open circuit See Long Description for complete error. Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Wed Mar 21 08:25:44 2007 Host : Jedi Ermes # : 260118185 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Axial Scan: 50103/8/6 (50-0204)Filament permanent open circuit. Issued by the heater board when an open has been detected and 3 restarts have been tried without success within 100 ms. ^C Z6,Z7= 97h,0h
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5128609-8 (replaces 2201185-3 - P9230TP) - HV Cable Cathode 2.3M - Used on Jedi HP 60 generator See Long Description for Details 5128609-8 Cathode HV Cable (2.3 meters long. Used on LightSpeed HP60 and BrightSpeed Systems) Part #: 5128609-8 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLE FRU: Yes Status: Orderable Comments: **FE ASSIST** 2300mm long HV cable from Cathode to HV tank Negative (212001261 May 4 2011) added per to NPI#019477 by sso#305010201 in Mar 28, 2010 Old FRU Compatible No Comments added per to ECR2109209 by sso#305024364 in Nov 2, 2010 Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5128609-8 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Direct ECO Number ECR2109209 2201185-3 - P9230TP - HV Cable Cathode 2.3M - Used on Jedi HP 60 generator, Model # 2371333-2 Part #: 2201185-3 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9230TP HV CABLE CATHODE 2.3M FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.