No water flow because of a vendor chiller problem.
Symptom
On HDE system Emergency Stop EMO trips 10 seconds after reset
Error message
2266420 Cabinet Monitor:A water flow failure was detected on the cabinet cooling system. Error: 2266420 Host: GE_MR2 Process: NSP File: SCP:CabMonHWO Process: NSP File: SCP:CabMonHWO Line: 569 Exception Class: Unknown Severity: Unknown Function: Power/Support Systems: Environmental Sensors/Controls PSD: FSE-XL Coil: 8CTL456 Scan: 841/6 Cabinet Monitor:A water flow failure was detected on the cabinet cooling system. The system goes to Emergency Stop State. Please contact your Service Representative and check cooling system with them.
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Area
- MR