GE Healthcare · CTLightSpeedVCT290014004290009001290009009
LightSpeed 2.X (Plus, QX/i) - GRE
Symptom
"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors
Error message
Error: 290014004 Unable to verify DARC system startup - scan disk array is non-recoverable Fri Dec 8 14:42:12 2006 Error: 290014004 Host: kghcateqxi Process: gre-verify-darc File: gre-verify-darc.cpp Line: 179 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - scan disk array is non-recoverable (exit status: 22) Error: 290009001The scanfile does not have enough views to make requested images. See LD Error: 290009001The scanfile does not have enough views to make requested. Tue Dec 6 11:07:48 2005 Error: 290009001 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 2684 The scanfile does not have enough views to make requested images. exam/series/scan = 60/2/1 expected number of images = 12 created number of images = 0 max number of images = 12 scan completed = 1 Error: 290009009 Image failed due to: IG error: NOT_ENOUGH_VIEWS Fri Jan 18 06:03:34 2008 Error:290009009 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 6076 Function: Image Generation : SBC Processing Image failed due to: IG error: NOT_ENOUGH_VIEWS. Number of retry so far: 1. IG node: 1
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5114772-3 DARC2 Dual Disk used ONLY on 4 slice SDAS See LD for more info 5114772-3 DARC2 Dual Disk used ONLY on 4 slice SDAS Part #: 5114772-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: DARC Computer - Dual Disk SDAS FRU: Yes Status: Orderable Comments: Added per GOC4 RSPL, ms13452 050304
- Area
- CT
Full cause & step-by-step resolution