GE Healthcare · MRExcite2252306222466622523572252293
Multicoil Bias Short or Open Faults Due to Cable
Symptom
Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.
Error message
2252306 TR Driver Fault Detected Short Circuit Source of Fault is Head TR See Long Description for Extended Error. 2252306 ERROR: EM_DCB_TR_SHORT_CKT TR Driver Fault Detected - Short Circuit. Source of Fault is %b. Value: %d mA Limit: %d mA Extended Error: If this error happens repeatedly then check for a short-circuited PIN diode or other short-circuit condition somewhere in the failing transmit chain. Use TR DD MC System Path diagnostic or Gather Load Data and verify that the short circuit values are less than 30mA. If error happens intermittently and cannot be traced to failed hardware then it may be due to gradient noise in the system. 2224666 Auto Prescan failed, Driver Module Error Occurred. Wed Apr 21 20:00:24 2004 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1334 Auto Prescan failed, Driver Module error occurred. 2252357 Driver Module error occurred. 2252357 1 lx-mr NSP APS:aps_fastta 135 Driver Module error occurred. 2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error mes
Applies to
- Models
- Excite
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2380635-5 Cable for 2380635-2 3.0T Quad Knee Coil See Long Description 2380635-5 Cable for 2380635-2 3.0T Quad Knee Coil
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.