Pular para o conteúdo principal
GE Healthcare · MROptima2267889

Gradient errors 2267889 and 2267904 due to Z axis XGD XPS.Will Not Scan with Gradient Errors.Replace XGD Power Supply XP…

Gradient errors 2267889 and 2267904 due to Z axis XGD XPS.Will Not Scan with Gradient Errors.Replace XGD Power Supply XPS 5148804-2 wqr.Optima MR450W.2267889 The Z axis XPS is reporting a High Voltage…

Symptom

Will Not Scan with Gradient Errors.

Error message

2267889 The Z axis XPS is reporting a High Voltage I-Bus Over-current Fault See Long Description Thu Apr 16 08:42:00 2009 Error: 2267889 Host: MR3T Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: sprlio Coil: HEAD Scan: 193/7 The Z axis XPS is reporting a High Voltage I-Bus Over-current Fault.-> Extended Error: Ensure that the XPS and XGA are properly connected, and that the XGA is operating properly by swapping XPS modules if necessary. If fault persists replace the failing XPS

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5148804-2 XGD Power Supply XPS 5148804-2 XGD Power Supply XPS ** 2nd Opinion Part ** Confirm appropriate extended error messages have been read and gradient diagnostics have been executed before placing order for this part. 212002950 May 28, 2009. Promoted this part for DVMR 3.0T NPI program, 212035293, 10 April 2008.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.