GE Healthcare · CTLightSpeedHiSpeed1831212257550013
LightSpeed 4.X - GRE
Symptom
Scanner Hardware Stopped Scan with open filament error in log. Scanner Hardware Stopped Scan with open filament error
Error message
Error: 183121 FIL/MA fault detected: Open filament circuit. See the long descripton for the full error code Tue Oct 11 13:12:27 2005 Error: 183121 Host: OBC Process: Generator Function: XRAY GENERATION : Filament Selection Scan: 22575/2/1 Type: Helical Exception Level: Pri/Most Severe FIL/MA fault detected: Open filament circuit. Desired focal spot = 0. (0 = large, 1 = small) <== Large Filament is open OR Fri Sep 22 07:03:11 2006 Error: 183121 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 341 Function: XRAY GENERATION : Filament Selection Scan: 50013/10/40 Type: Axial Exception Level: Pri/Most Severe Time: 07:03:11:045 Log Series: 393 FIL/MA fault detected: Open filament circuit. Desired focal spot = 1. (0 = large, 1 = small)<-------- Small
Applies to
- Models
- LightSpeed, HiSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2205390 CATHODE HV Tank used on 1.X, 2.x, 3.x, 4.x and HiSpeed LightSpeed Systems - Replaces 2161306 See Long Description for FEMC information and picture 2205390 CATHODE HV Tank used on 1.X, 2.x, 3.x, 4.x and HiSpeed LightSpeed Systems - Replaces 2161306 Part #: 2205390 Part Type: GP REPAIRABLES Part Weight: 30.06 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: BLOC HT CT FAST - POUR EPO FRU: Yes Comments: FLS - HV TANK CT FAST (-) CATHODE REPLACES 2161306 HDQ: 5 IN TRANSIT FROM GEMSE 4/3/98. REPLACES 2161306. PER G.ORME, ONLY DIFFERENCE IS PACKAGING, SO T3# REMAINS THE SAME. C.BRICKLER 4/6/98 ****NOTES FOR CD TEAM************************* ALL PARTS ON HAND HAVE BEEN REPAIRED USING THE NEW PROCEDURES. DO NOT USE THE 2205390H2 PART NUMBER ANYMORE.MS26561 DSM 11/16/01 ** FE ASSIST ** The HV adapter/Spacer part number 2171866 Kit that is included with the HV tanks But it is only used on older Highspeed Gantry due to the smaller bolt size. Refer to MI&CT ESERVICE NOTE SN07091114a for additional details. (212001261 21-Jan-2009) *** 2ND OPINION NOTES *** Check Oil in wells. Prior to ordering tube INSTALL NEW TUBE should be run to regenerate calseed values and set up the gencal table. The tube should be isolated and the KV loop test run to verify with 90% degree of certainty the tube is or is not at fault 4/7/98 2205390 1.0 0.5 SEC/REV CATHODE HV SUPPLY
- Area
- CT
Full cause & step-by-step resolution