GE Healthcare · MRSigna
Corduroy artifact due to rear pedestal cover loose and vibrating.
Symptom
MRIQ Corduroy Artifact See images in long description. Corduroy artifact was intermittently appearing on all types of scans with different coils. White pixels were showing up in raw data and spikes were seen on the scope (see attachment for Spike Noise Troubleshooting on Profile). Attachment also applies to Contour.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR