Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000107150042260001007260001022

LightSpeed 4.X - GRE

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260001071 The DCB has reset unexpectedly. See LD for more details Wed Nov 22 05:45:25 2006 Error: 260001071 Host: OBC Process: HOP File: hop_node_init.c Line: 347 Function: No System Function Reported Scan: 50042/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 05:45:24:285 Log Series: 85 The DCB has reset unexpectedly. The OBC has successfully reconnected to sub-system. Reason for reset as reported by CPU: External signal source. CPU reset register: 0x80 Node status: 0x382 State flags: 0x9800e82

Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the DCB See LD for more details Wed Nov 22 05:44:35 2006 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 50042/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 05:44:35:070 Log Series: 80 CAN serial link error. An CC error was detected when sending a message to the DCB. Message code: 1004 Node ID: 51 Object ID: 0 Send status: 5

Error: 260001022 Communications between the DCB sub-system and OBC have been lost NOTE: Error: 260001022 can also state CCB to OBC comm problem but is a different error in PSDB! See the long 

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170042P22 30A, 600V PANEL-MOUNTED FUSEHOLDER - used on GDAS See Long Discription for Details Part #: 46-170042P22 Part Type: GP CONSUMABLES Part Weight: 0.29 LBR Repairable: No Returnable: No Hazardous Material: N Description: 30A, 600V PANEL-MOUNTED FUSEHOLDER. FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required