Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260111603

LightSpeed 5.X (60kW)

Symptom

Step and Shoot (SmartStep or SmartView) hand controller not functioning See LD for more Step and Shoot (SmartStep or SmartView) hand controller not functioning When customer attempts to use the hand controller for VCT step and Shoot, the system will not initialize into step and shoot mode.

Error message

Error: 260111603 IHHC watchdog timeout. See long description for a full sample error. Wed Jul 2 11:44:37 2008 Error: 260111603 Host: tgp Process: tgp subsys File: hc wdog.c Line: 230 Wed Jul 2 11:44:36 2008 Host : Table Gantry Processor Ermes # : 260111603 Exception Class : Abort Severity : Primary File : hc wdog.c Line# : 230 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 IHHC watchdog timeout.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2199947 CT SONATA HAND HELD CONTROLLER CT SONATA HAND HELD CONTROLLER Part number 2199947 CT SONATA HAND HELD CONTROLLER Part number 2199947
Area
CT

Full cause & step-by-step resolution