Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2300150355120052300150241127205412230015013

LightSpeed 4.X - GOC2

Symptom

Fastcals failed to complete. Fastcals failed to complete

Error message

Error: 230015035 Cal_db failure See LD for complete error Tue Oct 18 07:20:43 2005 Error: 230015035 Host: ctls3 Process: scanmgr File: cal_db_file_io.c Line: 762 Cal_db failure, error number = 12, reason: Not enough space Tue Sep 20 08:36:512005 Error: 230015035 Host: ls16 Process: scanmgr File: cal_db_file_io.c Line: 231 Cal_db failure, error number = 98, reason: Address already in use

Error: 230015024 A corrupt file has been found in the cal database See LD for complete error Tue Sep 20 08:36:51 2005 Error: 230015024 Host: ls16 Process: scanmgr File: cal_db_file_io.c Line: 1983 Function: CAL Processing : SBC Processing A corrupt file has been found in the cal database. The corrupt file has been removed from the cal database and saved for later analysis. Corrupt File: /usr/g/cal_db/kv120/large_sfov/16x0625/large_spot/dasgain5/cal.1.acal Saved File: /usr/g/cal_db/recovery/kv120.large_sfov.16x0625.large_spot.das.cal.1.acal.1127205412

Error: 230015013 Failure opening a file for reading in the Cal Database. See LD for complete error Tue Sep 20 2005 08:36:512005 Error:230015013 Host: ls16 Process:scanmgr File: cal_db_file_io.c Line: 226 Function: CAL Processing : SBC Processing Failu

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) ConsoleClass C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2394897) 1104L.3_ H4.2CLASSC LightSpeed/ HiSpeed QXi Class C Software M4 5111370 5111371 LightSpeed 4.X Applications Software 405L.8_H4.2M4 FW11-04 FMI 25338
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues