Error 2274200 Port detection check failed due to defective RFCB board in RF HUB.Fails TPS Reset when booting up to Appli…

Error 2274200 Port detection check failed due to defective RFCB board in RF HUB.Fails TPS Reset when booting up to Applications or when Resetting TPS.Replace the RF Hub RF Control Board RFCB,wqr.Optim…

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

Error 2274200 Port detection check failed .The RF Hub hardware detected a port connector, but could not identify it Error 2274200 Port detection check failed .The RF Hub hardware detected a port connector, but could not identify it. Check cabling and port connectors. Port: Connector P2/1.5T/16 Channels Connector Fault Voltage: 930 mV Expected Voltage Range: 1634 mV - 1755 mV->

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5250098 1.5T RF Hub Control Board