Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Pink Pop-up Scan Disk Array has encountered a hard drive failure. If another hard drive fails you will lose scan data. See Long Description: Pink Pop-up message is displaced to the customer on boot up. Scan Disk Array has encountered a hard drive failure. If another hard drive fails you will lose scan data. "Scan Disk Array has encountered a hard drive failure. The System is functional, but if another hard drive fails you will lose scan data. Please Contact GE Service to have the disk array repaired as soon as possible."

Error message

Logical Drive ALERT : CHL:XX ID:X Drive Failure See long description Scan RAID event log ----200, 2011/08/18 15:23:09>: ID:6AE7CB75 Logical Drive ALERT: CHL:2 ID:0 Drive Failure

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5308760 Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible - used On GOC6 Disk Array See Long description: Part #: 5308760 Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible used on GOC6 Disk array ( 5308759 ). Part #: 5308760 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Seagate - PN ST32050310NS - Baracuda ES SATA Family Hard Drives with Carrier Compatible with 5308759 FRU: Yes Status: Orderable Comments: Per HDVCT RSPL, 212008898, 7/15/08. ** FE ASSIST ** This is a single Hard Drive used in the High Speed Disk Array on the GOC6 Console CT 750 HD and VCT (212001261 23-Jan-2009) ST32050310NS = 250 GB SATA II Hard Disk Drive
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required