ATOM CT - ( Advanced Technology On Mobile Gantry ) (ACT) 5.x HP60 LightSpeed 16 CT with Innova 4100 Advantix LCA Plus / LCV Plus Vascular System
Symptom
Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .
Error message
Error: 244540 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state See Long Description for complete error Sun Sep 26 08:36:46 2004 Error: 244540 4 000ctge04 fwmgr FwMgrEventHandler.c 1517 Running diagnostic: Scan Data Path Functional Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state Error: 244634 Subsystem State; DCB:down, CCB:down, TCB:down, JEDI:down, TGP/ORP:down, All Down See Long Description for complete error Thu Mar 18 21:29:36 2010 Error: 244634 Host: sim1 Process: fwmgr File: fwmgr utils.c Line: 664 Running: System Reset Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:down CCB:down TCB:down JEDI:down TGP/ORP:down
Applies to
- Models
- LightSpeed, Innova
- Product
- ATOM CT - ( Advanced Technology On Mobile Gantry ) (ACT) 5.x HP60 LightSpeed 16 CT with Innova 4100 Advantix LCA Plus / LCV Plus Vascular System See Long Description for details ATOM CT - ( Advanced Technology On Mobile Gantry ) (ACT) 5.x HP60 LightSpeed 16 CT with Innova 4100 Advantix LCA Plus / LCV Plus Vascular System What’s the ATOM Project? Clinical Efficacy Application Hepatocellular Carcinoma Hepatic Cancer for interventional procedure as TAE Neuroradiology Neuro intervention Vertebroplasty Pre-intervention Efficiency Decrease the operation time for intervention Prevention of infection inside the House less Co-medical Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5197802 Cherokee 5kW 48vdc power supply with fuse board. USED ON VCT gantry rotating frame. Supercedes previous Pioneer part numbers 5197802, 5183361, 5154556, and 5126259 Part #: 5197802 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCTHD 48V PS with FRU packaging FRU: Yes Status: Orderable Comments: **FE ASSIST** 5197802-2 is used on a HD-CT750 system, Do not use on a VCT system.(212001261 9/30/2010) .Per HDVCT RSPL, 212008898, 11/6/08 Part #: 5197802 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT Pioneer 5kW Power Supply Assembly Collector, with FRU packaging material FRU: Yes Status: Orderable Comments: Per ECR 2037621, 212008898, 1/5/07. If no stock, please order part # 5183361 --Description: VCT 5kW Power Supply Assembly, with ferrite and external shield. Part number change up history information: P/N 5197802 Cherokee " supply has two main differences: 1) The newer Cherokee supply is approximately 1.2 kg (2.6 lbs) Heavier then the older Pioneer Supply - this may cause a Gantry rebalance. 2) The large Ferrite bead is not visible on the Cherokee Supply because the ferrite is internal on the Cherokee Supply. P/N 5183361 was created to add an external shield that diverts the magnetic field energy away from the low channel DAS/Detector channels. The low channels on the VCT DAS/Detector are sensitive to magnetic fields that are produced from the power supply. This p/n replaced 5154556. P/N 5154556 was created to add a ferrite core to the 48 VDC output cable of the supply. This change was made to address streaking imaging artifacts that were sometimes seen with low mA scans by decreasing noise induced into the DAS/Detector. This p/n replaced 5126259. P/N 5126259 was the original VCT gantry rotating frame 48 vdc power supply. 5183361 48 vdc power supply with ferrite core and external shield 5154556 48 vdc power supply with ferrite core 5126259 48 vdc power supply without ferrite core or magnetic shield
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.