Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230000067500041680001500023000006450003

LightSpeed 4.X - GOC2

Symptom

"Scan hardware Stopped Scan" with Taxi data violation errors. "Scan hardware Stopped Scan" with Taxi data violation errors

Error message

Error: 230000067 Unexpected event occurred during scan data transfer See the long description for the full error code Sat Apr 17 20:34:13 2004 Error: 230000067 Host: darc Process: dataacq File: dip_thread.cxx Line: 390 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Xray-on Static Scan: 50004/4/1 Scan Seq Id: 168000 Unexpected event occurred during scan data transfer. Event bit: 7 Count: 1 Description: Global bus error: read attempted from the Test Data FIFO Scan aborted. OR Tue Mar 6 07:13:50 2007 Error: 230000067 Host: darc Process: dataacq File: vdip.cxx Line: 817 Function: Data Acquisition : Scan Data Acquisition Scan Type: Cine Scan: 50004/6/1 Scan Seq Id: 15000 Unexpected event occurred during scan data transfer. Event bit: 22 Count: 1 Description: DMA interrupt was not cleared Scan aborted.

Error: 230000064 TAXI data violation detected by forward error correction See the long description for the full error code Sat Apr 17 20:26:43 2004 Error: 230000064 Host: darc Process: dataacq File: dip_thread.cxx Line: 411 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2259530 H3 PMC DIP See LD for more info 2259530 H3 PMC DIP Part #: 2259530 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: H3 PMC DIP FRU: Yes Status: Orderable Comments: FLS - H3 DIP BOARD. USED ON WARP3. SUPPLIER IS SCI. SEE FE ASSIST. COLLECTION OF DEFECTIVES START7/23/01 MS5277 4/25/02- UNDER REPAIR DEVELOPMENT PER P REILLY. MS12464 5/7/03 - CHANGED BACK TO R0010BY P. REILLY MS12575 GELS - ***************FE ASSIST***************************** OLCE / PDB 01/11/28 ---H1 O2LESS --- QUARTZ FREQ. 44 MHZ : REF FRU : # 2216666,PCB H3 DIP
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.