LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. See long description for pop up message picture. A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. After clicking OK Fast Cal completes. ATTENTION: The system detected a problem with the DAS/Detector sub-system. Image Quality will be compromised. Scanning will NOT be disabled, but it is critical that the user call GE SERVICE to rectify the problem. Press 'OK" to proceed at the risk of Image Artifacts
Error message
Error: 230007000 Exclusion Test - Row 48:16B, Channel 552, Module 35 Open Channel* FAILED Z-AXIS EXCLUSION See LD for complete error. Tue Jul 19 03:54:27 2005 Error: 230007000 Host: ct01 Process: badchanmgr File: BadChannelMap.cxx Line: 533 Exclusion Test - Row 48:16B, Channel 552, Module 35 Open Channel* FAILED Z-AXIS EXCLUSION CRITERIA *. Error: 230007002 Exclusion Test - Row xx:xxx, Channel xxx, Module xx Open Channel* FAILED PR OXIMITY CRITERIA * See LD for a sample of entire error message. Fri Oct 28 09:32:03 2005 Error: 230007002 Host: presbyplano Process: badchanmgr File: BadChannelMap.cxx Line: 638 Exclusion Test - Row 64:32B, Channel 695, Module 44 Open Channel* FAILED PR OXIMITY CRITERIA *. Error: 230007001 Exclusion Test - Row xx:xxx, Channel xxx, Module xx Open Channel* FAILED X-AXIS EXCLUSION CRITERIA Mon Oct 3 11:30:58 2005 Error: 230007001 Host: ctm1 Process: badchanmgr File: BadChannelMap.cxx Line: 544 Exclusion Test - Row xx:xxx, Channel xxx, Module xx Open Channel* FAILED X-AXIS EXCLUSION CRITERIA *. Error: 290012010 Wrong corr type for det XX row XX correction type 0 (BAD_DET_NO_CORR) Sun Apr 12 13:37:20 2009 Error: 290012010 ig1 image_generation BadDet.cpp 29
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5202049 VCT 64 OUTER DETECTOR MODULE 5202049 VCT 64 OUTER DETECTOR MODULE - Part #: 5202049 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VCT64 OUTER DETECTOR MODULE - CHINA FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.