Symptom

Scanning aborts when doing 160 and 200mA, but works fine on lower mA stations.

Error message

Error : Jedi 60-0504 Inverter Gate Power Supply error. gate supply voltage has dropped below the level required to drive the IGBTs properly. see long description error code: 60-0504h message/explanation: inverter gate power supply error gate supply voltage has dropped below the level required to drive the igbts properly potential cause: 1. inverter (gate command board) 2. kv control board 3. generator input voltage too low or line impedance too high recommended action/troubleshooting guide: run hv power diagnostics

Applies to

Models
HiSpeed
Product
CT/e Dual, CT ProSpeed AII/FII Irix or Linux HiSpeed CT/e Dual CT Scanner System see long Description for sales information Product Name: CT/e Dual CT Scanner System Irix A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) N/A Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5126997 AC/DC mid power assy (replaces 2214995) See long description for details Part "2214995" changed to "5126997 Part #: 5126997 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: AC-DC 3PH MID POWER ASSY_FRU FRU: Yes Status: Orderable Comments: Not Available Part #: 2214995 Part Type: GP CONSUMABLES Part Weight: 8.99 LBR Repairable: No Returnable: No Hazardous Material: N Description: AC/DC 3PH MID POWER ASSY FRU: Yes Status: Orderable ====================== Parts Offered as: Current: 2214995 Harvested: Not Available Exchange: Not Available Comments: FLS - AC/DC 3PH MID POWER ASM U/O YMS PROSPEED LXI,DXI,FXI REPLACES 2193680 JEDI GENERATOR BRIDGE RECTIFIER=2121569 F1 FUSE : 2261172 ALSO USE ON CT/E GELS - DEFECTIVE PARTS TO BE PUT ON SHELVES FOR ANALYSIS PDB00/07/27 KANBAN CSS F00. 29-JUN-2000),POWER SUPPLY AC/DC 3PH MID
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.