GE Healthcare · MRSigna225210622238492223880

2252106 Failed to determine hardware state for AGP Due to Both MGD power supplies.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2252106 Failed to determine hardware state for AGP board within 20 seconds Error: 2252106 Host: rmns Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for AGP board within 20 seconds The board may be busy loading vxWorks or may be hung.

2252106 Failed to determine hardware state for SCP board within 20 seconds

2223849 AGP, SCP, APS EPROM power-up boot failed

2223880 Reset TPS was not successful. Board Failure Status APS Fail, SCP Fail, AGP Fail, VRE Pass. ErrorNo 9 See Long Description Error: 2223880 Host: OUMCGEMR1ET Process: NSP File: TardisReset.cc Line: 491 Exception Class: Unknown Function: Host: NSP Reset TPS was not successful. Board Failure Status: APS: Fail, SCP: Fail, AGP: Fail, VRE: Pass. ErrorNo: 9.-> Extended Error: The MGD failed to start a TPS Reset. Either the serial connection to the MGD STIF Board is not working or the Ethernet connection from the Host to the MGD Hub is not working. Run the APS, AGP, and SCP Host Data path ping tests to check the Ethernet connections. Run the APS, AGP, and SCP Board Level Diagnostics to check the SCSI Serial Expansion. The STIF Board serial connection is not tested. If both

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5181495-3 400W Quad Output 6Ux8HP cPCI Power Supply This is a replacement for EOPL 2294300-8