GE Healthcare · MRExciteSigna22482592248217284717422186302218501

2248259 Low or no Signal Strength Seen From Exciter 0 (0=Master,1=Slave) Due to UTNS3.

Symptom

The Download Of A Prescribed Series Failed

Error message

2248259 VERY LOW SIGNAL STRENGTH IN RRF CAL. Low or no Signal Strength Seen From Exciter 0 (0=Master,1=Slave) Mon May 3 13:52:03 2004 Error: 2248259 Host: lx-mr Process: NSP File: APS:RrfCalibrator::excCal Line: 2541 VERY LOW SIGNAL STRENGTH IN RRF CAL Low or no signal strength seen from Exciter 0 (0=Master,1=Slave) during RRF cal.

2248217 Receiver calibration error(s). Receiver DB out of spec. See Long Description 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APS Receiver calibration error(s). Receiver DB out of spec. R1 values that failed: Rx1: 12 10 Rx2: 12 11 10 9 Rx3: 12 11 10 9 Rx4: 12 11 10 Rx5: 12 11 10 9 Rx6: 12 11 10 9 Rx7: 12 11 10 9 Rx8: 12 10

2218630 The download of a prescribed series failed. 2218630 Host: Andys3T Process: scn File: RxManagerIF.m Line: 556 The download of a prescribed series failed. Internal status code is 2218501.

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2366218-2 UTNS3 See Long Description 2366218-2 UTNS3 EMC ED2 Compliant .
Area
MR

Full cause & step-by-step resolution