GE Healthcare · CT2600023585072326000102350011260003117260002318
Lightspeed 1.X - Octane
Symptom
Image artifact streaks. With no errors in message log. Occurs on long exams Abdomens (Not Heads), Streak artifact frequency increases, as gantry gets hot. See Long Description for images and details Image artifact streaks. With no errors in message log. Occurs on long exams Abdomens (Not Heads), Streak artifact frequency increases, as gantry gets hot.
Error message
Error: 260002358 Scan ABORT -- DAS Has Detected An Error. Scan Setup Errors Prevent The Data Acquisition System From Being Able To Scan. (See Previous Log Entries.) See Long Description for error example Tue Dec 7 07:23:22 2004 Error: 260002358 Host: DAS Process: DCB Firmware File: scanrx.c Line: 677 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 50723/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:23:23:210 Log Series: 37 Scan ABORT -- DAS Has Detected An Error. Scan Setup Errors Prevent The Data Acquisition System From Being Able To Scan. ( See Previous Log Entries .) NOTE: This is a status error message with many causes. Look for previous errors in the log. Do Not Use - use one of the other two Error: 260001023 See Long Description for complete error NOTE: The error message code included in this error is variable. You will need to refer to the log to find the cause of this error. Tue Dec 7 07:23:22 2004 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 367 Function: No System Function Reported Scan: 50723/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 07:23:22:910 Log Series: 37 The DCB has notified the system of an error condition whic
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2218158 SDAS +/-5VDC ANALOG PWR SUPPLY W/BRACKET. used on 1.x with H1 gantry See Long Description for details Part #: 2218158 Part Type: GP CONSUMABLES Part Weight: 1.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: DAS 5VDC ANALOG PWR SUPPLY W/BRACKET FRU: Yes Status: Orderable Comments: FLS - DAS +/-5VDC ANALOG PWR SUPPLY W/BRACKET U/O HELIOS DAS REPLACES 46-251198P46 MS08992 10/5/98 GENCO INDUSTRIES-WAUKESHA,WI PER FE - 46-251198P46 IS NOT ENTIRELY COMPATIBLE - 3 CONNECTION POINTS FOR WIRES - NEED 4- CANNOT SOLDER WIRE IN CORRECTLY DUE TO HOLE IN SOLDER POINT BEING TOO SMALL FOR MORE THAN 1 WIRE. TW/01/03
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.