GE Healthcare · MRDiscovery22676113577157
Error 2267611 The RF Hub reports that there is no LO signal on the cable from the exciter/clock ref module to the RF Hub…
Error 2267611 The RF Hub reports that there is no LO signal on the cable from the exciter/clock ref module to the RF Hub control board Due to the 3.0T RF Hub Control Board
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2267611 The RF Hub reports that there is no LO signal on the cable from the exciter/clock ref module to the RF Hub control board See Long Description Error - Tue Jul 14 02:32:39 2009 Error: 2267611 Host: C-3TMRI2 Process: diagCoord File: rfHub.cpp Line: 638 Diagnostic: RF Calib Iteration: 2 of infinite Action: 1 Task ID: 0x7f38530 Time: 3577157 Object: AGP Test Task Function: Software/Firmware/Config: N/A The RF Hub reports that there is no LO signal on the cable from the exciter/clock ref module to the RF Hub control board.-> Possible cause(s): NB DTx Module, RF Hub Control Board
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5250098-2 3.0T RF HUB CONTROL BOARD 5250098-2 3.0T RF HUB CONTROL BOARD. Comments: MUST run all receiver diagnostics, including Coil Datapath Diagnostic, prior to placing order for this part. 212002950 October 24, 2008. Part is being set up for 3.0T DVMR program RSPL. 501139061, 5-Mar-08.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.