Pular para o conteúdo principal
GE Healthcare · CTHiSpeed182119

System will not boot to or stay in applications.

Symptom

System will not boot to or stay in applications. System will not boot or stay in applications

Error message

Error: 182119 LAN Watchdog TIMEOUT for the SBC. See Long Description for more details Sun Sep 17 10:24:42 2006 Error: 182119 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Line: 556 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:24:20:560 Log Series: 37 LAN Watchdog TIMEOUT for the SBC. HW/SW Error: Did not receive LAN watchdog message.

Applies to

Models
HiSpeed
Product
HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2197229-5 ETHERNET SWITCH, 8 PORT used in 1X/2X/3X/4X GOC1/GOC2 See long description for more details Part #: 2197229-5 Part Type: GP CONSUMABLES Part Weight: 1.94 LBR Repairable: No Returnable: No Hazardous Material: N Description: ETHERNET SWITCH, 8 PORT FRU: Yes Status: Orderable Comments: FLS - ETHERNET SWITCH, 8 PORT. ADDEDPER PCN 220607. REPLACES P/N 2197229-3 AND -4. USE EXISTINGSTOCK FIRST. MS20544, 8/15/02.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.