GE Healthcare · CTLightSpeedVCT18221360046
LightSpeed 1.X - GRE
Symptom
Cradle will move a few inches and stop. Cradle will move in a few inches and stop
Error message
Error: 182213 Velocity tracking error See Long Description for the full error code Tue Mar 15 15:28:05 2005 Error: 182213 Host: ETC Process: Cradle File: CmdMotion.c 1.33 7/23/96 Line: 92 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:28:05:770 Log Series: 37 HW error: Velocity tracking error. Expected Velocity 500 (.001 mm/sec) Actual Velocity 9382 (.01 mm/sec) Current Position 60046 (.01 mm) Velocity Tolerance 250 (.001 mm/sec) Cradle Dir: 1 (0 = Out 1 = In [Gantry]).
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2269169 CRADLE DRIVE ASM FOR CATHAY TABLE. See Long Description for details Part #: 2269169 Part Type: GP REPAIRABLES Part Weight: 14.23 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: CRADLE DRIVE ASM FOR CATHAY TABLE FRU: Yes Status: Orderable Comments: FLS - CRADLE DRIVE ASM FOR CATHAY TABLE. P/N 2269169 REPLACES P/N 46-296235G2 PER PCN 220590USE FIRST EXISTING STOCK. MS20544 ON 8/23/02.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required