Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26010009350164

LightSpeed 5.X (60kW)

Symptom

Scan fails aborts with PDU HVDC Buss Fault Status Scan fails aborts with PDU HVDC Buss Fault Status.

Error message

Error: 260100093 PDU HVDC Buss Fault Status. The HVDC did not charge correctly in the alotted time Mon Sep 5 13:55:52 2005 Error: 260100093 Host: tgp Process: tgp_subsys File: scIsrService.c Line: 246 Running diagnostic: kV and mA Test: Xray Functional Test Mon Sep 5 13:55:51 2005 Host : Table Gantry Processor Ermes # : 260100093 Exception Class : Abort Severity : Primary File : scIsrService.c Line# : 246 Function : SYSTEM MONITORING : HVDC Monitor Scan Type : Static Scan: 50164/1/1 PDU HVDC Buss Fault Status. The HVDC did not charge correctly in the alotted time. Threshold 1: > 200V in 200ms Threshold 2: > 500V in 1.2s. Possible causes: HVDC Cable Short Circuit, SlipRing, or Jedi Inverter.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2113764-3 : Capacitors 6uF 370VAC (not 6NF) See long description for details 2113764-3 : Capacitors 6uF 370VAC (not 6NF) Part #: 2113764-3 Part Type: GP CONSUMABLES Part Weight: 1.10 LBR Repairable: No Returnable: No Hazardous Material: N Description: FILTER CAPS FRU: Yes Status: Orderable Parts Offered as: Current: 2113764-3 Harvested: Not Available Exchange: Not Available Comments: FLS - THERE IS A KNOWN PROBLEM WITH THE DESIGN AND REPLACING CAPS WILL DO IT AGAIN,FMI IS BEING WORKED ON,CONTACT CT SUPPORT BEFORE ORDERING PER JOHN SHUBERT 9-24-96 FILTER CAP C1,C2,C3,C7,C8,C9
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.