Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error:260107216 Axial Motor Control (AMC) Logic detected Gantry (Axial) VELOCITY Failure. see long description Mon Apr 3 09:38:04 2006 260107216 4 tgp tgp_subsys amc_mon.c 968 Mon Apr 3 09:38:05 2006 Host : Table Gantry Processor Ermes # : 260107216 Exception Class : Abort Severity : Primary File : amc_mon.c Line# : 968 Function : AXIAL CONTROL : Axial Control Scan Type : None/Unknown Scan: 50145/1/0 Axial Motor Control (AMC) Logic detected Gantry (Axial) VELOCITY Failure. Measured: 0.519 sec/rev Commanded: 0.500 sec/rev HW/SW Error: Retry Rx to verify probable gantry hardware problem.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2235342-3 AXIAL MOTOR W/ SPROCKET, HPOWER 2235342-3 AXIAL MOTOR W/ SPROCKET, HPOWER Part #: 2235342-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: AXIAL MOTOR W/ SPROCKET, HPOWER FRU: Yes Status: Orderable The axial motor FRU is item 1 in picture below:
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.