420V ACGD Breaker Trips Due To a Bad SGA Power Supply. 420 Volt ACGD Circuit Breaker On The PDU Is Tripped OFF.

Symptom

420 Volt ACGD Circuit Breaker On The PDU Is Tripped OFF Important safety issue. Do not have customer reset the 420 Volt ACGD breaker on the PDU cabinet.

Error message

2244587 (breaker tripped) The GP is reporting SGA Power Supply Power Off. Check 420V power breaker on PDU. If breaker is tripped, replace SGA-PS. Check that 420V power line is fully seated between PDU & SGA-PS. Push EMO Reset Switch on PDU. Else, replace SGA-PS.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009