GE Healthcare · CTDiscovery260134621260134617
Table does not respond to movement commands.
Symptom
Table does not respond to movement commands. Table does not respond to movement commands
Error message
Error: 260134621 Cradle absolute encoder warning See long description for full error Mon Aug 27 17:51:05 2007 Error: 260134621 Host: tgp Process: tgp subsys File: gtcb cmc drv req.cxx Line: 1002 Mon Aug 27 17:51:05 2007 Host : Table Sub-System Ermes # : 260134621 Exception Class : Secondary Severity : Pri/Soft File : gtcb cmc drv req.cxx Line# : 1002 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle absolute encoder warning. current pos = 70 um, temp = -18 degC Error : 260134617 Cradle clutch lines are disconnected. Cannot control cradle clutch See Long Description for complete error. Mon Nov 17 14:04:02 2008 Error: 260134617 Host: tgp Process: tgp_subsys File: gtcb_cmc_drv_req.cxx Line: 1040 Mon Nov 17 14:04:02 2008 Host : Table Sub-System Ermes # : 260134617 Exception Class : Secondary Severity : Pri/Soft File : gtcb_cmc_drv_req.cxx Line# : 1040 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle clutch lines are disconnected. Cannot control cradle clutch
Applies to
- Models
- Discovery
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5245271 GTCB3 (Global Table Control Board 3) This board replaces 5142177-3 GTCB2 and GTCB part # 5112470 thru 5112470-5, and 5142177. See long description for details Part #: 5245271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GTCB3 BOARD ASSY FRU: Yes Status: Orderable Comments: added per to ECO2080741 by sso#305010201 in May 8, 2009. Part will be available end of Sep, 2009, please order old p/n before new p/n on shelf.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.