GE Healthcare · CTBrightSpeed290009012290009016290010005

BrightSpeed HPower, Delight, Velocity or MERI HPower

Symptom

ASIR option installed and Recon shutting down The issue above was reported when installing ASIR upgrade on BrightSpeed MERI system with TIO console. FE installed GPU board, Recon Enchancement Option and ASIR Option. Recon would shutdown when booting up system. Only ASIR and Recon Enhancement option keys removed and system would boot up and scan.

Error message

Error: 290009012 Lost connection with IG(X) X =ig number that failed ie 1, 2 or 3 See the LD for the complete error code Error: 290009012 Lost connection with IG(X) X =ig number that failed ie 1, 2 or 3 Tue Apr 19 12:44:41 2005 Error: 290009012 Host: darc Process: recon_control File: recon_control_cb.cxx Line: 393 Lost connection with IG1 in fd: 33. Fri Oct 6 08:29:53 2006 Error: 290009012 Host: darc Process: recon_control File: recon_control_cb.cxx Line: 419 Socket event on 3, fd: 43.

Error: 290009016 Recon is going to SHUTDOWN mode due to XXXXXX. See Long Description for Error variation Tue Jul 12 14:31:49 2005 290009016 4 darc recon_control recon_control_cb.cxx 464 Recon is going to SHUTDOWN mode due to no IG can be used to perform reconstruction. Mon Aug 31 23:13:32 2009 Error: 290009016 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 6029 Recon is going to SHUTDOWN mode due to maximum number of consecutive suspends reached.

Error: 290010005 Socket error: socket is terminated because IG dies (error text varies) See Long Description for error examples. Tue Mar 16 10:40:09 2004 290010005 darc data_restore request_object.cxx 503 Socket error: socket is terminat

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
09BW35.11 For BrightSpeed Meri, Discovery CT590 RT and Optima CT580 See Long Description 09BW35.11 For BrightSpeed Meri, Discovery CT590 RT and Optima CT580 Note: Installed base systems with 09BW35.11 will be upgraded to 10BW06.5 by FMI after 10BW06.5 cut in manufacturing {ctuser@ct01}[2] check_config sw_hw_string: 09BW35.11_SP1-2-1.HP_P_ P16 _G_GMV P16 = Pancake Das sw_hw_string: BSD Elite: 09BW35.11_SP1-2-1.HP_P_P16_G_GMV BSD Elite Select: 09BW35.11_SP1-2-1.HP_S_P16_G_HLT {ctuser@anonymous}[1] swhwinfo -console TRUE_IN_ONE <-- Console type identification Software Collector Version 09BW35.11 p/n 5380493 (Orderable at Coakley) OS: 5372915 CTT 6.3.10 Application: 5372914 09BW35.11 Service Pack 1-2-1: 5393041 5372916 - Class C (Tab# 182) 5372917 - Class M (Tab# 182) Additional details: Discovery CT590 RT (16 slice CT for Oncology) Optima CT580 RT (8 slice CT for Oncology) Optima CT580 W (16 slice Radiology CT for Bariatric and Interventional, etc.)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.