Pular para o conteúdo principal
GE Healthcare · MRDiscovery22690572269055

Fails TPS Reset when booting up to Applications or when Resetting TPS due to IRF3.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2269057 The 1-Wire Network did not detect light on the DVMR Fiber Optic Link between the Exciter and the IRF3 Board. Error: 2269057 Host: andys3T Process: diagCoord File: oneWire.c Line: 1098 Task ID: 0x6db4420 Time: 4251 Object: AGP The 1-Wire Network did not detect light on the DVMR Fiber Optic Link between the Exciter and the IRF3 Board.

2269055 The 1-Wire Network did not detect communication on the DVMR Fiber Optic Link between the Exciter and the IRF3 Board.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5250066 IRF3 Board
Area
MR