GE Healthcare · CTLightSpeedVCT26000247050346
LightSpeed 2.X (Plus, QX/i) - GRE
Symptom
Cannot scan with Data Acquisition System Temperature is nearing Upper Limit
Error message
Error: 260002470 Highest temperature was seen is on Converter Board 41. See LD for more details Mon Apr 10 08:21:33 2006 Error: 260002470 Host: DAS Process: DCB Firmware File: ctemptst.c Line: 800 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50346/8/39 Type: Axial Exception Level: Pri/Soft Time: 08:21:33:105 Log Series: 32 WARNING: DAS Temperature is nearing the Upper Limit. Highest temperature was seen is on Converter Board 41. Measured: 55.5 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. Converter Board Bit Map: 0x010000000000. Bit Map Indicates Failed Board(s) for Multiple Board Errors. (Board Bit Map defines: BIT47 = CnvBrd 48 ... BIT0 = CnvBrd 1)
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5363182 (replaces 46-220248P3) H1-H2 Plenum Fan FRU GDAS Power Supply Fan FRU and SDAS See long description for details Part #: 5363182 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H1-H2 Plenum Fan FRU GDAS Power Supply Fan FRU FRU: Yes Status: Orderable Old FRU Compatible Yes Comments Per ECR 2089514, 212008898, 02 Nov 2009. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands No Use Demand History Yes Ultimate Prime 5363182 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2089514 46-220248P3 FAN: TUBE AXIAL, 115CFM, 115V, 60HZ, Used on H1 and H2 gantries for SDAS Fan / Plenum assembly Part #: 46-220248P3 Part Type: GP CONSUMABLES Part Weight: 1.50 LBR Repairable: No Returnable: No Hazardous Material: N Description: FAN: TUBE AXIAL, 115CFM, 115V, 60HZ FRU: Yes Status: Orderable Comments: FLS - BALL BEARING FAN NECESSARY FOR ROTATION ON CT GANTRY USED ON HELIOS 1 AND HELIOS 2 PER PCN 199761 MS08992
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required