GE Healthcare · MRDiscovery22675972267431
Cannot reset TPS with IRF3 not ready message due to Fiber optic transceiver (SFP) on RRX2.
Symptom
Cannot reset TPS with IRF3 not ready message Cannot reset TPS with IRF3 not ready message.
Error message
2267597 The RRX % module was not detected by IRF3. The system can not scan without this module See Long Description Mon Jun 13 10:58:22 2011 Error: 2267597 Host: OWS Process: diagCoord File: Irf3.cpp Line: 2593 Task ID: 0x488b060 Time: 9441 Object: AGP Exception Class: Notification Function: Software/Firmware/Config: N/A The RRX_2 module was not detected by IRF3. The system can not scan without this module.-> Extended Error : The system needs 1 RRX per 16 channels. Each RRX is enumerated through its connection to the RF Hub. A bad cable between a RRX and the RF-hub or a RF Hub switch board in the wrong slot can cause the RRX to appear to be configured for the wrong set of channels and thus not counted in the required number of RRX. If the RF Hub is not powered up, the RRX connected to it will all default to channels 113 - 128. 2267431 The IRF3 is not ready See Long Description Error: 2267431 Host: MR02 Process: NSPFile: AGP:Irf3RrfHWO Line: 1209Exception Class: Unknown Severity: UnknownFunction: Software/Firmware/Config: N/APSD: Unknown Coil: Unknown Scan: -1/-1 The IRF3 is not ready.-> Extended Error: This is just a warning message that indicates that another error message has al
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5183329-2 Fiber Optic Transceiver SFP 4Gb Multimode, RoHS-Compliant Note, as of March 24, 2010 this part can not be ordered yet. It will be a FRU. If you cant order the transceiver order the higher FRU.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.