Pular para o conteúdo principal
GE Healthcare · MRDiscovery22675972267431

Cannot reset TPS with IRF3 not ready message due to Fiber optic transceiver (SFP) on RRX2.

Symptom

Cannot reset TPS with IRF3 not ready message Cannot reset TPS with IRF3 not ready message.

Error message

2267597 The RRX % module was not detected by IRF3. The system can not scan without this module See Long Description Mon Jun 13 10:58:22 2011 Error: 2267597 Host: OWS Process: diagCoord File: Irf3.cpp Line: 2593 Task ID: 0x488b060 Time: 9441 Object: AGP Exception Class: Notification Function: Software/Firmware/Config: N/A The RRX_2 module was not detected by IRF3. The system can not scan without this module.-> Extended Error : The system needs 1 RRX per 16 channels. Each RRX is enumerated through its connection to the RF Hub. A bad cable between a RRX and the RF-hub or a RF Hub switch board in the wrong slot can cause the RRX to appear to be configured for the wrong set of channels and thus not counted in the required number of RRX. If the RF Hub is not powered up, the RRX connected to it will all default to channels 113 - 128.

2267431 The IRF3 is not ready See Long Description Error: 2267431 Host: MR02 Process: NSPFile: AGP:Irf3RrfHWO Line: 1209Exception Class: Unknown Severity: UnknownFunction: Software/Firmware/Config: N/APSD: Unknown Coil: Unknown Scan: -1/-1 The IRF3 is not ready.-> Extended Error: This is just a warning message that indicates that another error message has al

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5183329-2 Fiber Optic Transceiver SFP 4Gb Multimode, RoHS-Compliant Note, as of March 24, 2010 this part can not be ordered yet. It will be a FRU. If you cant order the transceiver order the higher FRU.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues