Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260107254260107251

LightSpeed RT16 & RT16 Xtra and Vision

Symptom

Gantry will not rotate or scan, alignment lights will not turn ON

Error message

Error: 260107254 Axial Motor Control (AMC) logic could not identify the AXIAL drive type. See Long Discription for Details Mon Jul 19 07:28:40 2010 Error: 260107254 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 241 Mon Jul 19 07:28:40 2010 Host : Table Gantry Processor Ermes # : 260107254 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 241 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 . Axial Motor Control (AMC) logic could not identify the AXIAL drive type. Reset scan hardware to retry.

Error: 260107251 Axial Motor Control (AMC) logic lost communication with the Axial Drive. See Long Discription for Details Error: 260107251 Axial Motor Control (AMC) logic lost communication with the Axial Drive. Mon Jul 19 07:41:43 2010 Error: 260107251 Host: tgp Process: tgp_subsys File: ds301.c Line: 828 Mon Jul 19 07:41:41 2010 Host : Table Gantry Processor Ermes # : 260107251 Exception Class : Secondary Severity : Sec/Soft File : ds301.c Line# : 828 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 . Axial Motor Control (AMC) logic lost communication with the AXIAL drive (Comm Watchdog Failure)

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P15 (250V,8A) Fuse used in axial dynamic brake circuit as CT2A1A10F1 and found in FAN ASSY- GANTRY, and F4, F5 on HEMRC Interface Board and other places See long description for details Part #: 46-170021P15 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 008.000A 0250V 3AB SLOW UL/ FRU: Yes Status: Orderable Comments: FLS - FUSE 8 AMP - 250 VOLT 3AB
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.