Pular para o conteúdo principal
GE Healthcare · MRSigna222466622245894062796

Auto Prescan Fails with TG Reached the High Limit

Symptom

Auto Prescan Fails with TG Reached the High Limit See error in Long Description 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan.

Error message

2224666 Auto Prescan failed, TG reached the high limit. Wed May 5 09:28:17 2004 Error: 2224666 Host: hpo1 Process: psc File: PscMainControl.cc Line: 1296 Auto Prescan failed, TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan.

2224589 TG reached the high limit. See Long Description Suite: MR01 Host: OC0 Proc: MSG_HANDLR0 Error: 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan. Note: On later systems, this same error code has been slightly reworded to read as follows: Thu Jun 12 11:06:12 2008 Error: 2224589 Host: MENEMR Process: NSP File: TPS:aps_exec Line: 4062796 TG reached the high limit. TG has been set to 190. Check patient positioning and if desired try Auto Prescan again or use Manual Prescan.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2293674-8 1.5T T/R Knee and Foot Array Coil Cable 2293674-8 1.5T T/R Knee and Foot Array Coil Cable
Area
MR

Full cause & step-by-step resolution